Abstract
In this paper, we demonstrated the comparison of digital-based CNT and analogue-based filament X-ray sources by taking the X-ray images of ACR mammo phantom for developing the intraoperative specimen X-ray system. The X-ray image of ACR mammo phantom taken at 25 kV 1 mA by filament shows overall the better image quality. However, when X-ray images of phantom were compared at 40 kV 1 mA by both sources, it showed that CNT X-ray source showed better image quality because of Aluminum window.
Original language | English |
---|---|
Title of host publication | Medical Imaging 2023 |
Subtitle of host publication | Physics of Medical Imaging |
Editors | Lifeng Yu, Rebecca Fahrig, John M. Sabol |
Publisher | SPIE |
ISBN (Electronic) | 9781510660311 |
DOIs | |
Publication status | Published - 2023 |
Event | Medical Imaging 2023: Physics of Medical Imaging - San Diego, United States Duration: 19 Feb 2023 → 23 Feb 2023 |
Publication series
Name | Progress in Biomedical Optics and Imaging - Proceedings of SPIE |
---|---|
Volume | 12463 |
ISSN (Print) | 1605-7422 |
Conference
Conference | Medical Imaging 2023: Physics of Medical Imaging |
---|---|
Country/Territory | United States |
City | San Diego |
Period | 19/02/23 → 23/02/23 |
Bibliographical note
Publisher Copyright:© COPYRIGHT SPIE. Downloading of the abstract is permitted for personal use only.
Keywords
- ACR mammo Phantom
- Carbon Nanotube X-ray source
- Digital X-ray source
- Filament X-ray source
- Intraoperative Specimen X-ray System