Skip to main navigation Skip to search Skip to main content

Clinical outcomes of magnesium-incorporated oxidised implants: A randomised double-blind clinical trial

  • Kang Mi Pang
  • , Jung Woo Lee
  • , Jin Yong Lee
  • , Jai Bong Lee
  • , Soung Min Kim
  • , Myung Jin Kim
  • , Jong Ho Lee

Research output: Contribution to journalArticlepeer-review

22 Citations (Scopus)

Abstract

Objectives: This study evaluated implant stability and clinical outcomes obtained with magnesium-incorporated oxidised implants (Mg titanate) and compared them to those blasted magnesium-incorporated oxidised implants (blasted Mg titanate). Patients and methods: Mg titanate was manufactured using the microarc oxidation (MAO) process. To obtain blasted Mg titanate, the MAO process was performed after blasting with TiO2 particles. The 15-month, randomised, double -blind clinical trial was conducted on 54 implants in 40 patients (Mg titanate, 27 implants in 18 subjects; blasted Mg titanate, 27 implants in 22 subjects), in whom 4.0 mm × 10 mm implants were placed to restore the unilateral loss of one or two molars in the mandible. The final prosthesis was attached 3 months postoperatively. Implant stability was measured by the implant stability quotient (ISQ) and periotest value (PTV) at the time of implant insertion, and 2, 3, and 15 months postoperatively. Marginal bone loss was evaluated at 2, 6, and 15 months postoperatively. Soft tissue analysis was performed at 15 months postoperatively. Results: Both implant systems showed high stability at all time points (>71). Mean marginal bone loss was 0.71 ± 0.65 mm and 0.75 ± 0.73 after 15 months in Mg titanate and blasted Mg titanate, respectively. There were no significant differences between the two implant surfaces with respect to ISQ(P = 0.988), PTV(P = 0.935), and marginal bone loss(P = 0.807) after 15 months. Conclusion: The success rate after 1 year of follow-up was 100% for both magnesium-incorporated oxidised implants. There were no significant differences in the clinical outcomes between the two surfaces at 15 months follow-up.

Original languageEnglish
Pages (from-to)616-621
Number of pages6
JournalClinical Oral Implants Research
Volume25
Issue number5
DOIs
Publication statusPublished - May 2014

Keywords

  • Bone-implant interactions
  • Clinical research
  • Clinical trials
  • Surface chemistry

Fingerprint

Dive into the research topics of 'Clinical outcomes of magnesium-incorporated oxidised implants: A randomised double-blind clinical trial'. Together they form a unique fingerprint.