Fluorescent intensity-based differential counting of FITC-doped silica nanoparticles: Applications of CD4+ T-cell detection in microchip-type flowcytometers

Hoyoung Yun, Hyunwoo Bang, Won Gu Lee, Hyunchang Lim, Junha Park, Joonmo Lee, Asif Riaz, Keunchang Cho, Chanil Chung, Dong Chul Han, Jun Keun Chang

Research output: Chapter in Book/Report/Conference proceedingConference contributionpeer-review

2 Citations (Scopus)

Abstract

Although CD4+ T-cells are an important target of HIV detection, there have been still major problems in making a diagnosis and monitoring in the third world and the region with few medical facilities. Then, it is necessary to use portable diagnosis devices at low cost when you put an enumeration of CD4+ T-cells. In general, the counting of CD4 below 200cells/μL makes it necessary to initiate antiretroviral treatment in adults (over 13 years old). However, lymphocyte subsets (including CD4 counts) of infants and young children are higher than those of adults. This fact shows the percentage of CD4+ T-cells of blood subsets, i.e., CD4/CD45%, CD4/CD8% or CD4/CD3% means a more reliable indicator of HIV infection than absolute counts in children. To know the percentage of CD4+ T-cell by using two fluorescent dyes of different emission wavelength, at least, one laser and two PMT detectors are in general needed. Then, it is so hard to develop a portable device like a 'toaster size' because this makes such a device more complex including many peripheral modules. In this study, we developed a novel technique to control the intensity of fluorescent dye-doped silica nanoparticles. I synthesized FITC-doped silica nanoparticles conjugated CD4 antibody 10 times brighter than FITC-conjugated CD45 antibody. With the difference of intensity of two fluorescent dyes, we measured two parameters by using only a single detector and laser. Most experiments were achieved with μFACS (microfabricated fluorescence-activated cell sorter) on an inverted microscope (IX71, Olympus). In conclusion, this method enables us to discriminate the difference between CD4 and CD4S in an intensity domain simultaneously. Furthermore, this technique would make it possible develop much cheaper and smaller devices which can count the number of CD4 T-cells.

Original languageEnglish
Title of host publicationBiomedical Applications of Micro- and Nanoengineering III
DOIs
Publication statusPublished - 2007
EventBiomedical Applications of Micro- and Nanoengineering III - Adelaide, Australia
Duration: 11 Dec 200613 Dec 2006

Publication series

NameProceedings of SPIE - The International Society for Optical Engineering
Volume6416
ISSN (Print)0277-786X

Conference

ConferenceBiomedical Applications of Micro- and Nanoengineering III
Country/TerritoryAustralia
CityAdelaide
Period11/12/0613/12/06

Keywords

  • CD4+ T-cell counting
  • Differential counting
  • Fluorescent dye-doped silica nanoparticles
  • Microchip-type flowcytometer

Fingerprint

Dive into the research topics of 'Fluorescent intensity-based differential counting of FITC-doped silica nanoparticles: Applications of CD4+ T-cell detection in microchip-type flowcytometers'. Together they form a unique fingerprint.

Cite this