Abstract
In this study, the weighted mass of standard molecular weight compounds (SMWC) was adopted to improve the accuracy of molecular weight distribution measured by the gel permeation chroma-tography (GPC). To evaluate the impact of SMWC compositions, the mixture prepared by constant, linear-weighted, and polynomial-weighted SMWC (3 g/L total) was used for the calibration of GPC, and results were confirmed with typical membrane-based size-exclusion experiments. The results obtained by GPC analysis revealed that the composition of SMWC in the feed significantly altered the final molecular weight cut-off (MWCO) of same membranes, and the polynomial-weighted SMWC provided the best match with the result of MWCO measured by the single-compound rejection experiment due to the enhanced signal intensity at the higher molecular weight compounds (>20,000 Da). Consequently, the preparation of SMWC in the polynomial manner should be suggested during the calibration and MWCO measurements of membranes by GPC.
| Original language | English |
|---|---|
| Pages (from-to) | 74-79 |
| Number of pages | 6 |
| Journal | Desalination and Water Treatment |
| Volume | 180 |
| DOIs | |
| Publication status | Published - Mar 2020 |
Bibliographical note
Publisher Copyright:© 2020 Desalination Publications. All rights reserved.
Keywords
- Gel permeation chromatography (GPC)
- Molecular weight cut-off (MWCO)
- Polyethylene glycols (PEGs)
- Ultrafiltration (UF) membrane
Fingerprint
Dive into the research topics of 'Preparation method of standard molecules for the precise estimation of molecular weight cut-off of membranes by gel permeation chromatography'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver